Bayer HealthCare Rapidlab 800 User Manual
Page 253

3Ć81
Maintaining the System
Table 3Ć1.āăSensor Fill Solutions
Sensor
Fill Solution
System
pH
pH
840, 850, 860
Na
+
Na
+
, K
+
,
Cl
–
,
Ca
++
850, 860
K
+
Na
+
, K
+
,
Cl
–
,
Ca
++
850, 860
Cl
–
Na
+
, K
+
,
Cl
–
,
Ca
++
850, 860
Ca
++
Na
+
, K
+
,
Cl
–
,
Ca
++
850, 860
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1.
Stop the system from the Menu screen:
a. Select
2
Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2.
Remove the sensor as shown in Figure 3-56:
a. Push up the latches on the measurement module door and lift the door.
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the sensor and pull the sensor up and out of the
measurement module.
Figure 3Ć56.āăRemoving a Measurement Sensor from an 850
Grasp the tab on the sensor and pull the
sensor up and out of the measurement
module
Push the springĆloaded latch to
the right
2
Menu Code
7
!