Bayer HealthCare Rapidlab 800 User Manual

Page 253

background image

3Ć81

Maintaining the System

Table 3Ć1.āăSensor Fill Solutions

Sensor

Fill Solution

System

pH

pH

840, 850, 860

Na

+

Na

+

, K

+

,

Cl

,

Ca

++

850, 860

K

+

Na

+

, K

+

,

Cl

,

Ca

++

850, 860

Cl

Na

+

, K

+

,

Cl

,

Ca

++

850, 860

Ca

++

Na

+

, K

+

,

Cl

,

Ca

++

850, 860

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

2.

Remove the sensor as shown in Figure 3-56:

a. Push up the latches on the measurement module door and lift the door.

b. Push the spring-loaded latch to the right.

c. Grasp the tab on the sensor and pull the sensor up and out of the

measurement module.

Figure 3Ć56.āăRemoving a Measurement Sensor from an 850

Grasp the tab on the sensor and pull the

sensor up and out of the measurement

module

Push the springĆloaded latch to

the right

2

Menu Code

7

!