Bayer HealthCare Rapidlab 800 User Manual
Page 565

Performance Characteristics
EĆ39
Level
Recovery
Expected
Observed
WRSD
*
n
Mode
3
Syringe
24
1.19
84.9
85.6
99.2
Capillary
24
0.61
85.4
85.6
99.7
Microcapillary
24
0.80
84.8
85.6
99.0
Microsyringe
24
1.08
85.2
85.6
99.5
Expired Gas
24
0.09
85.4
85.6
99.8
4
Syringe
24
1.34
150.0
149.7
100.2
Capillary
24
0.87
150.0
149.7
100.2
Microcapillary
24
2.04
150.1
149.7
100.2
Microsyringe
24
0.95
150.6
149.7
100.6
Expired Gas
24
0.19
150.0
149.7
100.2
5
Syringe
24
1.52
378.8
377.9
100.2
Capillary
24
1.55
378.7
377.9
100.2
Microcapillary
24
4.69
378.9
377.9
100.3
Microsyringe
24
2.35
379.6
377.9
100.5
Expired Gas
24
0.39
377.5
377.9
99.9
* WRSD = within-run standard deviation
Table EĆ34.āă860 Recovery and Precision TestingĊNa
+
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
119.4
Syringe
24
0.30
119.8
120.2
99.7
Capillary
24
1.33
120.0
119.8
100.2
Microcapillary
36
0.76
116.3
117.7
98.8
Microsyringe
24
0.47
119.4
120.3
99.2
(Continued)