System Administration
5Ć59
Table 5Ć23.āăCalibration Gas Values
Gas
Cal Range
Cal Default
Slope Range
Slope Default
pCO
2
2.00 – 7.90%
5.00%
(Cal CO
2
+ 2.9) –
99.9%
10.00%
pO
2
5.00 – 98.00%
12.00%
0.00% (fixed)
0.00%
1.
Access the Calibration Gas Values screen from the Menu screen:
a. Select
5
Operating Setup
and press
Enter
.
b. Select
6
Calibration Setup
and press
Enter
.
c. Select
3
Gas Values
and press
Enter
.
The Calibration Gas Values screen appears, as shown in Figure 5-36.
Figure 5Ć36.ā Calibration Gas Values Screen
You cannot change
pO
2
slope.
Press Clear Entry to
delete text in a field.
2.
Move the cursor to the field that you want to change.
3.
Type the new gas value and press
Enter
.
4.
Repeat steps 2 and 3 for each gas value you want to change.
5.
Press
Done
.
6.
Define another setup function or press
Exit Menu
to return to the Ready
screen.