Sample entry test – Bayer HealthCare Rapidlab 800 User Manual
Page 353
4Ć67
Troubleshooting the System
4.
Test the fluid detectors with whole blood:
a. Select
Whole Blood
and press
Enter
.
b. Insert a sample device containing a whole blood sample into the sample
port.
c. Press
Start Test
.
d. Remove the sample device when prompted.
e. Check the screen for the message, FD1, FD1A, FD2, FD5 acceptable, to
verify correct fluid detector function.
The system automatically performs a full wash sequence.
5.
Repeat step 4 if you want to test the fluid detectors with another type of fluid.
6.
Press
Exit Test
.
7.
Press
Exit Menu
.
Sample Entry Test
Use this procedure to test the ability of the system to:
S move the sample door and the sample probe
S identify the position of the sample door and the sample probe
S detect the type of sample device
S detect an obstruction
S calibrate the sample door to various types of sample devices
1.
Access the Sample Entry Test from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
1 Fluidics System
and press
Enter
.
c. Select
6 Sample Entry
and press
Enter
.
The Sample Entry Test screen appears as shown in Figure 4-9.
6
1
Menu Code
3