Bayer HealthCare Rapidlab 800 User Manual
Page 550

EĆ24
800 Series Operator's Manual
Table EĆ19.āă850 Recovery and Precision TestingĊNa
+
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
24
0.25
115.6
115.5
100.1
Capillary
24
0.29
115.6
115.5
100.1
Microcapillary
24
0.47
115.1
115.1
100.0
Microsyringe
24
0.42
115.4
115.3
100.1
pH/Lytes
24
0.43
121.0
120.9
100.1
2
Syringe
24
0.49
150.9
151.0
99.9
Capillary
24
0.46
150.9
151.2
99.8
Microcapillary
24
0.80
151.2
151.1
100.0
Microsyringe
24
0.49
151.0
151.3
99.8
pH/Lytes
24
0.75
151.3
151.6
99.7
3
Syringe
24
0.39
174.2
174.2
100.0
Capillary
24
0.39
174.3
174.1
100.1
Microcapillary
24
1.81
173.9
173.9
100.0
Microsyringe
24
0.45
174.1
173.9
100.1
pH/Lytes
24
1.33
170.4
170.2
100.2
* WRSD = within-run standard deviation