EĆ48
800 Series Operator's Manual
Substance
Level of Interference
Interferes with . . .*
Fetal Hemoglobin at 80%
tHb
COHb
MetHb
HHb
–0.9 g/dL
+3.1%
–1.1%
–2.6%
Carboxymethylcellulose
tHB
COHb
O
2
Hb
HHb
–2.0 g/dL
+10.0%
–10.0%
+6.5%
* Analytes that are within interference criteria stated in Table EĆ40 are not listed.
Precision on Controls
Quality control materials were analyzed on the 800 series CO-ox module. The
results are presented here.
Precision on aqueous quality control materials was estimated using four 800
systems. One run per instrument was made over a eight day period. Three replicates
of each control level were analyzed in each run.
Table E-43 summarizes the results of the 800 CO-ox module precision on QC
materials.
Table EĆ43.āă800 COĆox Module Precision Results
Parameter
Level
n
Mean
WRSD
*
TotSD
[
tHb
1
96
9.36
0.05
0.25
2
96
13.98
0.07
0.37
3
96
21.60
0.08
0.54
FO
2
Hb
1
96
5.79
0.11
0.28
2
96
13.56
0.18
0.30
3
96
22.01
0.22
0.30
(Continued)