3Ć25
Maintaining the System
11. Connect the sample tubing to the sample tubing connector on the measurement
module.
12. Press
Continue
.
A wash sequence starts. When the wash sequence sequence finishes, a
message box appears prompting you to perform a two-point calibration.
13. Press
Yes
to perform a two-point calibration.