Standby and shutdown, Placing the system in standby – Bayer HealthCare Rapidlab 800 User Manual

Page 493

background image

System Administration

5Ć83

Standby and Shutdown

This section describes procedures to place the 800 system in standby and to shut
down the system.

Standby is an inactive state that disables the automatic calibration functions to
reduce reagent consumption. During Standby, the system continues to perform
purge sequences to maintain the integrity of the sensors, and it performs the Auto
Clean sequence.

Shutdown discontinues all operations in an orderly fashion. Always perform a
shutdown procedure before you disconnect the power cord to perform maintenance
or troubleshooting procedures.

Placing the System in Standby

Place the system in standby when sample analysis is not required for a prolonged
period of time. Using standby reduces reagent consumption by turning off
automatic calibrations. While in standby, the system performs the scheduled auto
clean and purges the sensors on a regular basis.

1.

Access the Set Standby screen from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

1 Standby

and press

Enter

.

2.

Place the system in standby.

If you want to . . .

Then . . .

schedule a date and time to
exit standby

Type the date and time.

a. Type the month, date, and year in the selected

format and press

Enter

.

b. Type the hour and minutes in the format

HH:MM. Enter 0 – 23 hours and 0 – 59
minutes.

c. Press

Done

to enter standby.

place the system in standby
indefinitely

Press

Done

.

The Standby screen appears. If you do not type the date and time to exit
standby, the Date and Time fields do not appear.

The system remains in standby until the scheduled date and time or until you
exit standby manually.

7

Menu Code

1