3Ć52
Rapidlab 800 Operator's Manual
Figure 3Ć33.āăRemoving a Roller Cage
Platen
Roller Cage
Roller Cage Shaft
Large Tubing
Small Tubing
Tubing Cuff
WARNING
Wear safety glasses, gloves, and a laboratory coat when handling
bleach.
6.
Clean the roller cage:
a. Clean the rollers with a lint-free tissue moistened with a 10% solution of
household bleach.
b. Clean the interior surface of the roller cage with a lint-free swab moistened
with the bleach solution.
c. Clean the roller cage shaft with a lint-free tissue moistened with the bleach
solution.
d. Rinse the rollers, the cage, and the shaft with reagent water.
e. Dry the roller cage thoroughly.
f. Ensure that the rollers turn freely.
7.
Reinstall the roller cage:
a. Replace the roller cage on the shaft.
b. Turn the roller cage on the shaft until it stops.
c. Press the roller cage down until the cage snaps into place.
8.
Connect the tubing as described in
Replacing the Pump Tubing
, page 3-39.