Defining automatic transmission of results – Bayer HealthCare Rapidlab 800 User Manual
Page 448

800 Series Operator's Manual
5Ć38
Defining Automatic Transmission of Results
Use this procedure to select automatic transmission of patient sample results
directly to an LIS, HIS, or data management system. You can use this option only
if your 800 system is connected to an LIS or data management system. Refer to
Selecting Automatic Transmission of QC Results, page 5-55, for the procedure to
define auto send for QC results.
Table 5Ć12.āăAuto Send Patient Sample Options
Option
Description
On
The system automatically sends patient sample results to an LIS or data
management system. The default setting is On.
Off
Results are not automatically transmitted. The operator must press
Send
to
transmit results or press
Do Not Send
to prevent transmission of results.
1.
Access the Systems Options screen from the Menu screen:
a. Select
6 System Setup
and press
Enter
.
b. Select
5 Systems Options
and press
Enter
.
The System Options screen appears, as shown in Figure 5-23.
Figure 5Ć23.ā System Options Screen for an 850
Move to the Auto Send
frame and select
Patient.
2.
Move to the Auto Send frame.
3.
Select
Patient
and press
Enter
.
A dark box indicates the option is on.
4.
Press
Done
.
5.
You can define another setup function or press
Exit Menu
to return to the
Ready screen.
6
Menu Code
5