Pump flow rate test – Bayer HealthCare Rapidlab 800 User Manual
Page 343

4Ć57
Troubleshooting the System
5.
Perform the appropriate action.
If you want to . . .
Then . . .
test another reagent
repeat steps 2 through 4.
return to the Menu
screen
press
Exit Test
.
6.
Press
Exit Menu
.
Pump Flow Rate Test
Use this procedure to test the flow rate of the reagent pump and to calibrate the
reagent pump, if necessary.
1.
Access the Pump Flow Rate Test from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
1 Fluidics System
and press
Enter
.
c. Select
2 Pump Flow Rate
and press
Enter
.
The Pump Flow Rate Test screen appears as shown in Figure 4-3.
Figure 4Ć3.ā Pump Flow Rate Test Screen
The flow rate units are mL/sec.
Check here for results.
2.
Press
Start Test
.
2
1
Menu Code
3