D75 lamp failure – Bayer HealthCare Rapidlab 800 User Manual
Page 331

4Ć47
Troubleshooting the System
D75 Lamp Failure
Problem
The CO-ox module detects lamp levels that are inadequate for sample analysis.
Solutions
Perform the Lamp On/Off Test as described in CO-ox Optics Test, page 4-75.
Remove the lamp cover from the CO-ox module and visually verify that the
lamp is on. If the lamp is not on, replace it as described in Replacing the CO-ox
Lamp in Section 3.
Visually verify that light is passing through the CO-ox sample chamber.
If the D code reappears, contact your Service Representative.
D75