Selecting Automatic Acceptance of QC Results
Use this function if you are connected to an LIS or data management system and
you want the 800 system to automatically accept QC results. When you use Auto
Accept with Auto Send, the 800 system accepts all QC samples and sends them to
the connected LIS or data management system. You can review the QC results at
the LIS or data management system and, if required, change the status to reject or
discard.
NOTE:
Auto Send must be turned on to have results automatically sent to an LIS
or data management system.
Table 5Ć20.āăAuto Accept QC Options
Option
Description
On
The system automatically accepts QC results and, when Auto Send is on,
sends the results to an LIS or data management system.
Off
The operator accepts, rejects, or discards QC results at the end of analysis.
The default setting is Off.
1.
Access the Global QC Settings screen from the Menu screen:
a. Select
5 Operating Setup
and press
Enter
.
b. Select
1 QC Setup
and press
Enter
.
c. Select
2 Global QC Settings
and press
Enter
.
The Global QC Settings screen appears, as shown in Figure 5-32.
Figure 5Ć32.ā Global QC Settings Screen
Move to the Auto
Accept QC frame and
select On or Off.
2.
Move to the Auto Accept QC frame.
3.
Select
On
or
Off
and press
Enter
.
4.
Press
Done
.
2
1
Menu Code
5