Bayer HealthCare Rapidlab 800 User Manual
Page 616

HĆ24
800 Series Operator's Manual
Figure HĆ6.āăRemoving the Sample Port
Drip Tray
Sample Door
Retainer Ring
Mount
Sample Port
O-ring
4.
Grasp the sample port and drip tray and pull it to the right to remove it.
NOTE:ĄĂ
The sample port and the drip tray are one piece.
Wear safety glasses, gloves, and a laboratory coat when
handling the reagents.
5.
Clean any deposits on the sample port, the drip tray, and the mount with a
lint-free swab moistened with a 10% solution of household bleach.
6.
Rinse the sample port, the drip tray, and the mount with reagent water.
NOTE:ĄĂ
Ensure that the three O-rings are in place.
7.
Reinstall the sample port, matching the tab on the sample port to the notch in
the retainer ring.
8.
Push the tab on the retainer ring away from you until it locks in place.
9.
Reinstall the sample door, ensuring that it snaps in place.
10. Press
Continue
.
11. Press
No
.
12. Press
Home
to return to the Ready screen.
Cleaning the Exterior Surfaces
Wear safety glasses, gloves, and a laboratory coat when handling
the reagents.
1.
Using a 10% solution of household bleach, wipe all exterior surfaces
including:
S the sample entry components and the drip tray
S the waste area
WARNING
WARNING