Selecting parameters for analysis – Bayer HealthCare Rapidlab 800 User Manual
Page 422

5Ć12
800 Series Operator's Manual
1.
Access the Select Default Panels screen from the Menu screen:
a. Select
6 System Setup
and press
Enter
.
b. Select
4 Panels
and press
Enter
.
c. Select
1 Default Panel
and press
Enter
.
The Select Default Panel screen appears, as shown in Figure 5-4.
Figure 5Ć4.ā Select Default Panel Screen for an 850 System
Select the panel you
want and press Enter.
2.
Select the appropriate panel and press
Enter
.
Custom panels are listed as Panels 1–5. Custom panels are defined during
system setup by the system administrator or authorized personnel. Refer to
Defining Custom Panels on page 5-9.
3.
Press
Done
when you finish.
Selecting Parameters for Analysis
Use this procedure to select the default set of parameters you want the system to
analyze. If you select a parameter to be off, the system does not measure or
calculate results during analysis, and the parameter name does not appear on the
Results screen or the printed report. The default value for all parameters is on,
except for O
2
SAT, O
2
CT, and [Qsp/Qt (est,T)].
NOTE:
Any calculated parameters that rely on another parameter value, which has
been selected to be off, are not calculated. However, the parameter name appears on
the Results screen, but no value is reported and the parameter name and value do
not appear on the printed report.
1.
Access the Parameters screen from the Menu screen:
a. Select
6 System Setup
and press
Enter
.
b. Select
3 Parameters
and press
Enter
.
The Parameters screen appears, as shown in Figure 5-5.
Menu Code
6
4
1
Menu Code
6
3