Analyzing capillary tube samples – Bayer HealthCare Rapidlab 800 User Manual
Page 106

2Ć8
Rapidlab 800 Operator's Manual
Analyzing Capillary Tube Samples
The following table lists the minimum volumes required by each system to analyze
capillary tube samples:
Table 2Ć2.āăMinimum Sample Volumes
System
Minimum Sample Volume
840
844, 845
0
90
µ
L
140
µ
L
850
854, 855
110
µ
L
160
µ
L
860
864, 865
125
µ
L
175
µ
L
NOTE:
If an 844, 845, 854, 855, 864, or 865 detects insufficient sample in the
measurement module, the system analyzes the sample as a microsample. If you
want the system to analyze a different set of parameters for the sample, press
Change Parameters to select another panel. Refer to Selecting Parameter Panels,
page 2-29, for more information.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
CAUTION:
Insert the fire-polished end of the capillary tube into the sample port to
prevent damage to the capillary seal.
1.
Prepare the sample and insert the fire-polished end of the capillary tube into
the sample port, as shown in Figure 2-3.
Figure 2Ć3.ā Inserting a Capillary Tube
Sample Port
Capillary Sample
!