Bayer HealthCare Rapidlab 800 User Manual
Page 603

Installation
HĆ11
e. Pull the paper from under the platen and push it through the slot in the
printer cover.
f. Push the printer lever down.
35. Close the printer cover:
a. Pull up the paper spool.
b. Insert the paper into the paper slot on the spool and turn three or four
rotations. Press
Paper Advance
if there is not enough paper.
c. Push down the spool.
36. Access the Menu screen and set up your system.
Refer to System Administration in Section 5 for more information about setting
up the system.
37. Initiate a prime sequence from the Menu screen to remove bubbles from the
reagent lines:
a. Select
2 Maintenance
and press
Enter
.
b. Select
3 Prime
and press
Enter
.
c. Select
All
and press
Enter
.
d. Press
Done
.
e. Watch the movement of wash solution to verify that the wash reagent flows
through the system during the wash sequence that automatically follows the
prime sequence.
38. Verify the gas flow rates for cal gas and slope gas:
a. Access the Valves Test from the Menu screen.
b. Select
Cal Gas
and press
Start Test
.
c. Insert an aspiration adapter into the sample port.
d. Immerse the open end of the aspiration adapter into a small container of
reagent water.
e. Press
Start Test
.
f. Verify that a steady stream of bubbles flows into the water.
g. Press
Stop Test
.
h. Verify that the bubbles stop flowing.
i. Repeat steps b through h for slope gas.
j. Remove the aspiration adapter.
k. Press
Exit Test
.
39. Access the Temperature test from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
2 Temperature
and press
Enter
.
c. Check the screen to verify that the sample temperature is 37.00
±
0.15
°
C.
2
Menu Code
3
3
Menu Code
2