Installation

HĆ11

e. Pull the paper from under the platen and push it through the slot in the

printer cover.

f. Push the printer lever down.

35. Close the printer cover:

a. Pull up the paper spool.

b. Insert the paper into the paper slot on the spool and turn three or four

rotations. Press

Paper Advance

if there is not enough paper.

c. Push down the spool.

36. Access the Menu screen and set up your system.

Refer to System Administration in Section 5 for more information about setting
up the system.

37. Initiate a prime sequence from the Menu screen to remove bubbles from the

reagent lines:

a. Select

2 Maintenance

and press

Enter

.

b. Select

3 Prime

and press

Enter

.

c. Select

All

and press

Enter

.

d. Press

Done

.

e. Watch the movement of wash solution to verify that the wash reagent flows

through the system during the wash sequence that automatically follows the
prime sequence.

38. Verify the gas flow rates for cal gas and slope gas:

a. Access the Valves Test from the Menu screen.

b. Select

Cal Gas

and press

Start Test

.

c. Insert an aspiration adapter into the sample port.

d. Immerse the open end of the aspiration adapter into a small container of

reagent water.

e. Press

Start Test

.

f. Verify that a steady stream of bubbles flows into the water.

g. Press

Stop Test

.

h. Verify that the bubbles stop flowing.

i. Repeat steps b through h for slope gas.

j. Remove the aspiration adapter.

k. Press

Exit Test

.

39. Access the Temperature test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

2 Temperature

and press

Enter

.

c. Check the screen to verify that the sample temperature is 37.00

±

0.15

°

C.

2

Menu Code

3

3

Menu Code

2