Stopping the system – Bayer HealthCare Rapidlab 800 User Manual
Page 233

3Ć63
Maintaining the System
Stopping the System
Use this procedure to stop the system when you perform maintenance activities
such as replacing components. Stopping the system discontinues all fluidic
activities such as calibrations and sample analyses.
CAUTION:
Because no fluids reach the sensors while the system is stopped,
stopping the system for a prolonged period of time may affect the performance of
the sensors.
1.
Stop the system from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
2.
Perform the required maintenance task.
3.
Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
4.
Take the appropriate action.
If . . .
Then . . .
you want to perform a
two-point calibration
press
Yes
.
you do not want to initiate a
two-point calibration
a. Press
No
.
b. Press
Home
to return to the Ready screen.
If calibrations were scheduled while the system was stopped, the system performs
the calibrations when you exit the stopped mode.
!
2
Menu Code
7
Procedural
Notes