Bayer HealthCare Rapidlab 800 User Manual
Page 107
2Ć9
Operating the System
CAUTION:ĄĂ
To ensure the accuracy of the CO-ox measurement, close the
CO-ox cover before pressing Analyze.
2.
Press
Analyze
.
If . . .
Then . . .
there is sufficient
sample volume
the system automatically aspirates the sample and moves it
to the measurement module.
the system prompts
you to move the
sample
a. Look to see if the sample has reached point A. If not,
turn the sample pump clockwise to move the leading
edge of the sample to point A, as shown in Figure 2-4.
b. Press
Sample in Place
.
Figure 2Ć4.ā Sample at Point A on an 850
Point A
Sample Pump
3.
When prompted, remove the sample device.
4.
Type the required information in the Patient Information screen and then press
Done
.
5.
Perform the appropriate task.
If . . .
Then . . .
the system is moving
the sample
the results appear on the screen and are updated until
analysis is complete. Continue with step 8.
you are moving the
sample
continue with step 6.
!