Selecting automatic qc file assignment – Bayer HealthCare Rapidlab 800 User Manual
Page 463

System Administration
5Ć53
Selecting Automatic QC File Assignment
Use this procedure to control the way QC results are assigned to a QC file.
Table 5-19 describes the options for assigning QC results.
Table 5Ć19.āăAuto Identify Options
Option
Description
On
The system automatically determines the appropriate QC file
to assign QC sample results, by comparing the QC sample
results for pH and pCO
2
to the ranges entered in setup for
these parameters.
NOTE:
If the pH, pCO
2
,
or tHb sensor is disabled or not in
calibration, Auto ID does not work.
Off
The operator assigns the QC sample results to the appropriate
file. The default value is Off.
1.
Access the Global QC Settings screen from the Menu screen:
a. Select
5 Operating Setup
and press
Enter
.
b. Select
1 QC Setup
and press
Enter
.
c. Select
2 Global QC Settings
and press
Enter
.
The Global QC Settings screen appears with the cursor in the Auto ID frame,
as shown in Figure 5-31.
Figure 5Ć31.ā Global QC Settings Screen
Select On or Off from
the Auto ID field and
press Enter.
2.
Select
On
or
Off
and press
Enter
.
3.
Press
Done
.
4.
Define another setup function or press
Exit Menu
to return to the Ready
screen.
2
1
Menu Code
5