Bayer HealthCare Rapidlab 800 User Manual
Page 555

Performance Characteristics
EĆ29
Substance
Concentration Tested
Thiocyanate
80 mg/dL
Bilirubin (Direct)
30 mg/dL
Bilirubin (Total)
34 mg/dL
Creatinine
30 mg/dL
Hydroxybutyrate
200 mg/dL
Lactate
100 mg/dL
Urea
500 mg/dL
Uric Acid
10 mg/dL
* Interference by dopamine and structurally related drugs is dependent on glucose concentration. However,
even at high glucose concentrations, therapeutic levels of dopamine do not interfere with the glucose
measurement.
The following anticoagulants were found not to interfere with glucose recovery at
the indicated concentrations but they cannot be used on the 860 system due to
potential interferences on other sensors or analytes.
Table EĆ24.āăPotential Interference on Other Analytes
Substance
Concentration Tested
Citrate
1000 mg/dL
Potassium Oxalate
1000 mg/dL
Refer to Sample Collection Devices and Anticoagulants in Section 1 for the specific
requirements on sample handling and anticoagulants.