Performance Characteristics

EĆ29

Substance

Concentration Tested

Thiocyanate

80 mg/dL

Bilirubin (Direct)

30 mg/dL

Bilirubin (Total)

34 mg/dL

Creatinine

30 mg/dL

Hydroxybutyrate

200 mg/dL

Lactate

100 mg/dL

Urea

500 mg/dL

Uric Acid

10 mg/dL

* Interference by dopamine and structurally related drugs is dependent on glucose concentration. However,

even at high glucose concentrations, therapeutic levels of dopamine do not interfere with the glucose
measurement.

The following anticoagulants were found not to interfere with glucose recovery at
the indicated concentrations but they cannot be used on the 860 system due to
potential interferences on other sensors or analytes.

Table EĆ24.āăPotential Interference on Other Analytes

Substance

Concentration Tested

Citrate

1000 mg/dL

Potassium Oxalate

1000 mg/dL

Refer to Sample Collection Devices and Anticoagulants in Section 1 for the specific
requirements on sample handling and anticoagulants.