Cleaning the sample path with bleach – Bayer HealthCare Rapidlab 800 User Manual
Page 229

3Ć59
Maintaining the System
8.
Perform a two-point calibration from the Menu screen:
NOTE:ĄĂ
Glutaraldehyde may cause excessive drift to the K
+
and Ca
++
sensors.
Repeat calibrations until the system performs a successful calibration.
a. Select
1 Calibration
and press
Enter
.
b. Select
2 Two-point
and press
Enter
.
9.
Analyze a minimum of two levels of quality control material to verify sensor
performance.
10. Empty and clean the waste bottle, waste outlets, and waste outlet cover as
described in
Emptying the Waste Bottle
, page 3-45.
Cleaning the Sample Path with Bleach
Materials required:
S 10% solution of household bleach
NOTE:
Dilute household bleach (5.25% sodium hypochlorite) 1:10 with
reagent-quality water.
S aspiration adapter
S lint-free tissue or swabs
S test/blank reference sensor (TB5)
S test/blank glucose sensor (TB4)
S test/blank lactate sensor (TB4)