Reviewing and printing workload statistics – Bayer HealthCare Rapidlab 800 User Manual
Page 234

3Ć64
Rapidlab 800 Operator's Manual
Reviewing and Printing Workload Statistics
The 800 system collects workload data and provides month-to-date and
year-to-date workload statistics reports. The workload statistics report includes the
total cycle count of all patient samples, calibrations, and QC samples performed by
the system. The yearly statistics are reset every January 1 after the system performs
the first sample analysis or calibration. The system automatically prints a workload
statistics report.
1.
Access the Workload Stats screen from the Menu screen:
a. Select
4 Data Recall
and press
Enter
.
b. Select
6 Workload Stats
and press
Enter
.
2.
Press
Print Statistics
to print a workload statistics report.
The workload statistics report is shown in Figure 3-36.
Figure 3Ć36.āăWorkload Statistics Report
WORKLOAD STATISTICS REPORT Apr 12 1994 11:42
System 850-1001
Total cycle cnt: 025356
Month-to-Date Statistics Apr 1 1994 to Apr 12 1994
Patient Samples: 000573
QC Samples: 000102
1 Point Cal: 000861
2 Point Cal: 000145
Year-to-Date Statistics Jan 1 1994 to Apr 12 1994
Patient Samples: 009150
QC Samples: 001675
1 Point Cal: 006680
2 Point Cal: 002196
Total cycle count is the
number of patient
samples,QC samples,
calibrations, and
operator-initiated washes.
Month-to-Date
Workload Statistics
Year-to-Date
Workload Statistics
3.
Press
Done
to return to the Menu screen.
4.
Press
Home
to return to the Ready screen.
4
Menu Code
6