Lactate biosensor interfering substances – Bayer HealthCare Rapidlab 800 User Manual

Page 557

background image

Performance Characteristics

EĆ31

Lactate Biosensor Interfering Substances

To test for interferences, serum was spiked with a potentially interfering substance
up to the test concentration shown in the following tables. The interference was
calculated by comparing the spiked sample to an unspiked sample immediately
preceding it. To evaluate the interference from fluoride/oxalate, the blood was
drawn in grey-top tubes and compared to the lactate recovery in heparinized blood.

The following table lists substances that were found not to interfere with the lactate
measurement. At the concentrations listed, these compounds produced less than
0.3 mmol/L (2.7 mg/dL) error in the recovered lactate concentration.

Table EĆ26.āăSubstances Showing No Detectable Interference

Substance

Concentration Tested

Chlorpromazine

17 mg/dL

Dopamine

1 mg/dL

Ethanol

350 mg/dL

Salicylate

50 mg/dL

Sodium Nitroprusside

70 mg/dL

Thiocyanate

80 mg/dL

Heparin

20,000 U/dL

Epinephrine

2 mg/dL

Norepinephrine

2 mg/dL

Phenobarbital

15 mg/dL

Glutamate

16 mg/dL

Hetastarch

30%

Acetoacetate

40 mg/dL

Ascorbate

8 mg/dL

Dilantin

14 mg/dL

(Continued)