Lactate biosensor interfering substances – Bayer HealthCare Rapidlab 800 User Manual
Page 557

Performance Characteristics
EĆ31
Lactate Biosensor Interfering Substances
To test for interferences, serum was spiked with a potentially interfering substance
up to the test concentration shown in the following tables. The interference was
calculated by comparing the spiked sample to an unspiked sample immediately
preceding it. To evaluate the interference from fluoride/oxalate, the blood was
drawn in grey-top tubes and compared to the lactate recovery in heparinized blood.
The following table lists substances that were found not to interfere with the lactate
measurement. At the concentrations listed, these compounds produced less than
0.3 mmol/L (2.7 mg/dL) error in the recovered lactate concentration.
Table EĆ26.āăSubstances Showing No Detectable Interference
Substance
Concentration Tested
Chlorpromazine
17 mg/dL
Dopamine
1 mg/dL
Ethanol
350 mg/dL
Salicylate
50 mg/dL
Sodium Nitroprusside
70 mg/dL
Thiocyanate
80 mg/dL
Heparin
20,000 U/dL
Epinephrine
2 mg/dL
Norepinephrine
2 mg/dL
Phenobarbital
15 mg/dL
Glutamate
16 mg/dL
Hetastarch
30%
Acetoacetate
40 mg/dL
Ascorbate
8 mg/dL
Dilantin
14 mg/dL
(Continued)