Bayer HealthCare Rapidlab 800 User Manual
Page 123

2Ć25
Operating the System
1.
Analyze the arterial, and the venous or mixed venous samples you want to use
for a-v studies.
2.
Select the first patient sample for a-v studies:
a. Select
Menu
.
b. Select
4 Data Recall
and press
Enter
.
c. Select
1 Patient Data
and press
Enter
.
The Patient Data Search Criteria screen appears.
3.
Enter the patient ID for the sample and press
Done
.
The Patient Data Search Log screen appears.
4.
Select the patient sample for a-v studies and press
Reporting Options.
The Reporting Options message box appears.
5.
Select
Select for aĆv Studies
and press
OK
.
The First a-v Sample Selected message box appears.
6.
Press
OK
.
If . . .
Then . . .
only one sample is found
the Patient Data Search Result screen
appears.
Press
Reporting Options
. The Reporting
Options message box appears.
1
more than one sample is found
the Patient Data Search Log screen appears.
a. Select the next sample for a-v studies
and press
Enter
.
b. Press
Reporting Options
. The
Reporting Options message box appears.
7.
Select
Select for aĆv Studies
and press
OK
.
The Confirm Samples for a-v Studies message box appears.
NOTE:ĄĂ
Record and use the sequence numbers to access the sample results
from the Patient Data Search Log.
8.
Enter the Qt value if you want VO
2
and DO
2
results.