Bayer HealthCare Rapidlab 800 User Manual
Page 548
EĆ22
800 Series Operator's Manual
Level
Recovery
Expected
Observed
WRSD
*
n
Mode
Capillary
24
1.20
71.7
71.3
100.6
Microcapillary
18
1.36
71.1
71.3
99.7
Microsyringe
18
0.55
70.5
71.3
98.9
Expired Gas
18
0.49
71.0
71.3
99.6
* WRSD = within-run standard deviation
Table EĆ18.āă850 Recovery and Precision TestingĊpO
2
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
21
0.16
28.2
28.5
98.9
Capillary
21
0.18
28.9
28.5
101.4
Microcapillary
18
0.35
29.0
28.5
101.8
Microsyringe
18
0.25
28.5
28.5
100.2
Expired Gas
18
0.16
27.9
28.5
98.0
2
Syringe
24
0.22
50.5
49.9
101.2
Capillary
18
0.34
49.2
49.9
98.6
Microcapillary
24
0.41
49.4
49.9
99.0
Microsyringe
24
0.13
49.8
49.9
99.8
Expired Gas
24
0.10
49.4
49.9
99.1
3
Syringe
18
0.34
85.3
85.6
99.6
Capillary
22
0.33
85.5
85.6
99.9
Microcapillary
18
0.99
84.5
85.6
98.7
Microsyringe
18
0.70
85.6
85.6
100.0
Expired Gas
18
0.20
84.7
85.6
98.9
(Continued)