Relocating your system – Bayer HealthCare Rapidlab 800 User Manual

Page 612

background image

HĆ20

800 Series Operator's Manual

2.

Plug the power cord into the AC power receptacle.

At startup, the system proceeds through a series of initial tests and then
displays

Initializing

in the banner. Any errors incurred during initialization are

posted to the status log.

3.

Access the Service Setup screen from the Menu screen:

a. Select

8

Service Setup

and press

Enter

.

b. Select

4 System ID

and press

Enter

.

4.

Type the service password and press

OK

.

5.

Select the system configuration(s) that you want and press

Enter

.

6.

Enter the CO-ox module serial number and press

Enter

.

7.

Press

Done

.

8.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

9.

Access the Menu screen and set up your system.

Refer to System Administration in Section 5 for more information about setting
up the system.

10. Initiate a two-point calibration from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

2 Two-point

and press

Enter

.

c. Verify acceptable drift.

11. Initiate a tHb slope from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

9 tHb Slope

and press

Enter

.

c. Verify acceptable drift.

12. Analyze QC material and verify acceptable results.

Relocating Your System

No special handling or preparation is required to relocate your system to another
area in your laboratory. To store your 800 system for an extended period of time or
to ship it to another location, perform the following procedures, or contact your
Bayer Diagnostics Service Representative for assistance.

Materials required:

S 10% solution of household bleach
S sterile water
S reagent water

8

Menu Code

4

1

Menu Code

2

1

Menu Code

9