Daily maintenance, Cleaning the exterior surfaces – Bayer HealthCare Rapidlab 800 User Manual
Page 176

3Ć8
Rapidlab 800 Operator's Manual
Daily Maintenance
The daily maintenance schedule is based on analyzing 30 samples per day, unless
otherwise noted. If your laboratory analyzes more than 30 samples per day,
perform daily maintenance more frequently.
Cleaning the Exterior Surfaces
Materials required:
S 10% solution of household bleach
NOTE:
Dilute household bleach (5.25% sodium hypochlorite) 1:10 with reagent
quality water.
S reagent-quality water
44
S lint-free tissues or swabs
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
S Wipe with a 10% solution of household bleach all exterior surfaces, including:
S the sample entry components and the drip tray
S the waste area