Bayer HealthCare Rapidlab 800 User Manual
Page 453
System Administration
5Ć43
1.
Access the Correlation Coefficients screen from the Menu screen:
a. Select
5 Operating Setup
and press
Enter
.
b. Select
8 Correlation
and press
Enter
.
The Correlation Coefficients screen appears, as shown in Figure 5-26.
Figure 5Ć26.ā Correlation Coefficients Screen for an 850
Press Clear Entry to
delete text in a field.
2.
Type the slope and offset values you want for each parameter.
3.
Press
Done
when you finish.
4.
Define another setup function or press
Exit Menu
to return to the Ready
screen.
If you press Done before completing all the fields, the Incomplete Data Entry
message appears. Press
OK
and complete the required fields.
If you type an invalid value in a field and press Done, the Invalid Entry message
box appears. Press
OK
and type a valid entry in the field.
pH values on this screen are always expressed in pH units and not in nmols/L.
5
Menu Code
8
Procedural
Notes