Reporting results with calibration drift – Bayer HealthCare Rapidlab 800 User Manual
Page 450

800 Series Operator's Manual
5Ć40
Figure 5Ć24.ā System Options Screen for an 850
Move to the Reporting
Resolution frame and
select High or Low.
2.
Select
High
or
Low
and press
Enter
.
3.
Press
Done
.
4.
Define another setup function or press
Exit Menu
to return to the Ready
screen.
The option that you select for the Reporting Resolution is saved with each patient
and QC sample. If you recall the sample data later and have changed the
resolution, the system prints the report using the resolution selected when the
sample was saved.
Reporting Results with Calibration Drift
Use this procedure to define the way the system processes patient sample results
when a sensor is out of calibration. When the system detects a sensor that has
excessive calibration drift, it cannot analyze samples until the drift is corrected. By
selecting Yes for Report Results with Cal Drift, you have the option to let the
system analyze a sample when a sensor is out of calibration.
Table 5Ć14.āăReport Results with Cal Drift Options
OptionDescription
Yes
The system analyzes the sample and logs the event in the status log. The
printed report contains the message, not in calibration.
No
A D2 code appears and the affected sensor is turned off. Analysis continues
but no results are generated for the affected sensor. The default value is No.
Procedural
Notes