Asćrequired procedures, Emptying the waste bottle – Bayer HealthCare Rapidlab 800 User Manual
Page 215

3Ć45
Maintaining the System
AsĆRequired Procedures
Emptying the Waste Bottle
Materials required:
S 10% solution of household bleach
NOTE:
Dilute household bleach (5.25% sodium hypochlorite) 1:10 with
reagent-quality water.
S reagent water
S lint-free tissue or swabs
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
WARNING
Wear safety glasses, gloves, and a laboratory coat when performing
this procedure.
1.
Press and hold down the waste bottle latch while removing the waste bottle
from the compartment, as shown in Figure 3-27.
Figure 3Ć27.āăRemoving the Waste Bottle
Waste Bottle
Waste Detector
Waste Outlet Cover
Waste Outlet
Press down the latch.
2.
Cover the waste bottle with the attached cap to avoid spills when transporting
the bottle.