Analyzing vacuum tube samples – Bayer HealthCare Rapidlab 800 User Manual
Page 116

2Ć18
Rapidlab 800 Operator's Manual
Analyzing Vacuum Tube Samples
The following table lists the minimum volumes required by each system to analyze
vacuum tube samples
:
Table 2Ć5.āăMinimum Sample Volumes
System
Minimum Sample Volume
840
844, 845
0
90
µ
L
140
µ
L
850
854, 855
110
µ
L
160
µ
L
860
864, 865
125
µ
L
175
µ
L
If you want the system to analyze a different set of parameters for the sample,
press Change Parameters to select another panel. Refer to Selecting Parameter
Panels, page 2-29, for more information.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1.
Prepare the sample and insert an aspiration adapter into the sample port.
2.
Insert the adapter into the vacuum tube, as shown in Figure 2-12.
Figure 2Ć12.ā Inserting an Aspiration Adapter with a Vacuum Tube Sample
Sample Port
Vacuum Tube
Aspiration Adapter