Bayer HealthCare Rapidlab 800 User Manual
Page 156

2Ć58
Rapidlab 800 Operator's Manual
Press
Cancel to interrupt a calibration.
The system starts a wash and the Wash screen appears. The system returns to the
Ready screen at the end of the wash.
The maximum time between tHb slope calibrations is 90 days. One day before a
tHb slope calibration is scheduled, a status message appears on the screen
indicating that a calibration is due.
Procedural
Notes