Bayer HealthCare Rapidlab 800 User Manual
Page 354
4Ć68
800 Series Operator's Manual
Figure 4Ć9.ā Sample Entry Test Screen
Perform the automatic
test to check the
sample door and
sample probe positions.
Check here for results.
2.
Test the positions of the sample door and the sample probe:
a. Select
Automatic
and press
Enter
.
b. Press
Start Test
.
c. Check the screen for the message, Sample entry module acceptable, to
verify that the sample door and sample probe function correctly.
d. Press
Exit Test
.
3.
Test the ability of the system to recognize the sample device:
a. Select
Manual
and press
Enter
.
b. Insert a sample device, with the plunger pulled back, into the sample port
when prompted.
c. Press
Start Test
.
d. Remove the sample device when prompted.
e. Check the screen to verify that the system correctly identifies the size and
the type of the sample device.
f. Press
Exit Test
.
4.
Test the ability of the sample probe to detect obstructions:
a. Select
Manual
and press
Enter
.
b. Push in completely the plunger of a syringe and insert the syringe into the
sample port when prompted.
c. Press
Start Test
.
d. Remove the syringe when prompted.
e. Check the screen to verify that the sample probe detects an obstruction and
identifies the size of the sample device.