Filling the measurement sensors – Bayer HealthCare Rapidlab 800 User Manual
Page 252

3Ć80
Rapidlab 800 Operator's Manual
Figure 3Ć55.āăFilling the KCl Reservoir
Electrode
Compartment
4M KCl Fill Solution
Fill Line
OĆring
OĆring
Sample Path
c. Tap the front face of the sensor with your knuckle to release any bubbles.
d. Gradually fill the KCl reservoir with KCl fill solution to the fill line.
CAUTION:ĄĂ
Do not overtighten the reservoir cap. Overtightening can damage
the gasket and cause leaks.
e. Reinstall the reservoir cap and hand-tighten.
11. Continue with
Reinstalling a Sensor
, page 3-89.
Filling the Measurement Sensors
Use this procedure to fill any measurement sensors that do not have sufficient fill
solution. Refer to Table 3-1 to determine the appropriate fill solution to use.
Materials required:
S appropriate fill solution
S lint-free tissue
NOTE:
If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
NOTE:
The pO
2
and pCO
2
sensors do not require fill solution.
NOTE:
The glucose and lactate biosensors do not require fill solution.
!
860