Bayer HealthCare Rapidlab 800 User Manual
Page 546

EĆ20
800 Series Operator's Manual
Table E-16 through Table E-22 summarize the results of the 850 system whole
blood, expired gas, and electrolyte recovery and precision testing.
Table EĆ16.āă850 Recovery and Precision TestingĊpH
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
24
0.002
7.194
7.193
100.0
Capillary
24
0.003
7.194
7.194
100.0
Microcapillary
24
0.003
7.170
7.175
99.9
Microsyringe
24
0.002
7.229
7.225
100.1
pH/Lytes
18
0.003
7.220
7.218
100.0
2
Syringe
24
0.001
7.399
7.405
99.9
Capillary
24
0.002
7.397
7.393
100.1
Microcapillary
24
0.003
7.402
7.396
100.1
Microsyringe
24
0.002
7.383
7.397
99.8
pH/Lytes
18
0.004
7.357
7.356
100.0
3
Syringe
24
0.004
7.493
7.493
100.0
Capillary
24
0.003
7.502
7.505
100.0
Microcapillary
24
0.006
7.449
7.459
99.9
Microsyringe
24
0.003
7.557
7.560
100.0
pH/Lytes
18
0.005
7.673
7.676
100.0
* WRSD = within-run standard deviation