Bayer HealthCare Rapidlab 800 User Manual
Page 144

2Ć46
Rapidlab 800 Operator's Manual
3.
Type the search criteria and press
Enter
after you complete each field.
If you want to
search . . .
Then . . .
for a selected date
type the date in both Analysis Date fields.
from earliest date to
the present date
leave the Analysis Date and Analysis Time fields blank.
from the earliest date
to a specific date
leave the Analysis Date From field blank, and type the
specified date in the Analysis Date To field.
from a specific date
to the present date
type the start date in the Analysis Date From field, and
type nothing in the Analysis Date To field.
for a specific range of
dates
type the start date in the Analysis Date From field, and
type the end date in the Analysis Date To field.
by shift or any other
specific time period
type the start time in the Analysis Time From field and the
end time in the Analysis Time To field.
4.
Press
Done
.
If . . .
Then . . .
more than one sample
is found
the QC Data Search Log screen appears as shown in
Figure 2-26.
Use the log to select reports you want to view, print, or
transmit.
one sample is found
the QC Data Search Result screen appears as shown in
Figure 2-27.
Figure 2Ć26.ā QC Data Search Log Screen
Press Reporting Options
to print or transmit a
report selected from the
log.