800 Series Operator's Manual
5Ć62
Configuring for External Devices
Use the following procedures to configure the 800 system for any of the following
external devices:
S 270 CO-oximeter
S 800 series compatible ticket printer
S Bayer Diagnostics data management systems
S bar code scanner
S line printer
S laboratory or hospital information system
Table 5-25 lists the ports you can use with each device.
Table 5Ć25.āăPorts and Devices
Select . . .
To connect . . .
Serial port 1
the ticket printer.
Serial port 2 and 3
the 270 CO-oximeter and an LIS or data management system.
Parallel port
other printer types, such as a line printer.
Barcode
the bar code scanner.
1.
Access the Port Selection screen from the Menu screen:
a. Select
6 System Setup
and press
Enter
.
b. Select
6 Communications
and press
Enter
.
The Port Selection screen appears, as shown in Figure 5-38.
Figure 5Ć38.ā Port Selection Screen for an 850
Select the port you
want to use.
6
Menu Code
6