Checking the solution levels in the sensors – Bayer HealthCare Rapidlab 800 User Manual
Page 185

3Ć17
Maintaining the System
Checking the Solution Levels in the Sensors
Use this procedure to ensure that the sensors contain fill solution to the levels
described in Figure 3-5.
Figure 3Ć5ā.ăFill Solution Levels
Sensor
Fill Solution Level
pH, K
+
, Cl
–
, and Ca
++
nearly full and with a bubble at the top
Na
+
to the top
Reference
to the fill line
NOTE:
The pO
2
and pCO
2
sensors do not require you to maintain them, even
though they have fill solution. Slight discoloration of the fill solution in the pO
2
and pCO
2
sensors is normal.
NOTE:
The glucose and lactate biosensors do not require fill solution.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1.
Stop the system from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
2.
Open the measurement module door as shown in Figure 3-6:
a. Push up the latches on the measurement module door.
b. Lift the door.