Bayer HealthCare Rapidlab 800 User Manual
Page 547
Performance Characteristics
EĆ21
Table EĆ17.āă850 Recovery and Precision TestingĊpCO
2
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
24
0.28
14.0
14.3
97.9
Capillary
24
0.08
14.7
14.3
102.5
Microcapillary
24
0.20
14.5
14.3
101.7
Microsyringe
24
0.14
14.1
14.3
98.7
Expired Gas
24
0.14
14.5
14.3
101.6
2
Syringe
21
0.12
21.3
21.4
99.7
Capillary
24
0.12
21.5
21.4
100.5
Microcapillary
18
0.59
20.7
21.4
96.9
Microsyringe
18
0.42
21.3
21.4
99.4
Expired Gas
18
0.77
21.0
21.4
98.3
3
Syringe
18
0.43
35.1
35.7
98.2
Capillary
24
0.25
36.4
35.7
102.0
Microcapillary
18
0.50
34.5
35.7
96.6
Microsyringe
18
0.34
35.9
35.7
100.7
Expired Gas
18
0.30
34.9
35.7
97.8
4
Syringe
24
0.72
50.5
49.9
101.3
Capillary
24
0.29
48.4
49.9
97.1
Microcapillary
24
0.47
51.1
49.9
102.4
Microsyringe
24
0.21
49.9
49.9
100.0
Expired Gas
24
0.26
49.9
49.9
100.1
5
Syringe
21
1.13
71.7
71.3
100.5
(Continued)