Performance Characteristics

EĆ21

Table EĆ17.āă850 Recovery and Precision TestingĊpCO

2

Level

Mode

n

WRSD

*

Observed

Expected

Recovery

1

Syringe

24

0.28

14.0

14.3

97.9

Capillary

24

0.08

14.7

14.3

102.5

Microcapillary

24

0.20

14.5

14.3

101.7

Microsyringe

24

0.14

14.1

14.3

98.7

Expired Gas

24

0.14

14.5

14.3

101.6

2

Syringe

21

0.12

21.3

21.4

99.7

Capillary

24

0.12

21.5

21.4

100.5

Microcapillary

18

0.59

20.7

21.4

96.9

Microsyringe

18

0.42

21.3

21.4

99.4

Expired Gas

18

0.77

21.0

21.4

98.3

3

Syringe

18

0.43

35.1

35.7

98.2

Capillary

24

0.25

36.4

35.7

102.0

Microcapillary

18

0.50

34.5

35.7

96.6

Microsyringe

18

0.34

35.9

35.7

100.7

Expired Gas

18

0.30

34.9

35.7

97.8

4

Syringe

24

0.72

50.5

49.9

101.3

Capillary

24

0.29

48.4

49.9

97.1

Microcapillary

24

0.47

51.1

49.9

102.4

Microsyringe

24

0.21

49.9

49.9

100.0

Expired Gas

24

0.26

49.9

49.9

100.1

5

Syringe

21

1.13

71.7

71.3

100.5

(Continued)