Bayer HealthCare Rapidlab 800 User Manual
Page 276
3Ć104
Rapidlab 800 Operator's Manual
b. Select
7 Stop System
and press
Enter
.
2.
Allow the lamp to cool for at least 5 minutes.
WARNING
Ensure that the lamp has been off for at least 5 minutes to allow
sufficient time for it to cool.
3.
Remove the lamp cover from the CO-ox module.
4.
Remove the lamp connector, as shown in Figure 3-71.
Figure 3Ć71ā.ăReplacing the COĆox Lamp
Lamp Cover
Lamp
COĆox Pump
5.
Remove the old lamp from the lamp housing and discard it.
6.
Place a new lamp in the housing aligning the lamp with the guide pin.
7.
Pinch the ends of the clamp together and replace it in the bracket.
8.
Install the lamp connector.
9.
Reinstall the lamp cover.
10. Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
11. Press
No
.
12. Calibrate the new lamp:
a. Select
3 Troubleshooting
and press
Enter
.
Menu Code
3
3
2