Removing obstructions from the coćox sample path – Bayer HealthCare Rapidlab 800 User Manual
Page 375

4Ć87
Troubleshooting the System
Removing Obstructions from the COĆox Sample Path
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
NOTE:
Do not use a needle on the syringe that you use to introduce reagent water
or air into the CO-ox sample tubing. Using a needle may damage the CO-ox
sample tubing.
1.
Stop the system from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
2.
Inspect the CO-ox sample tubing and waste tubing for obstructions, as shown
in Figure 4-20.
Figure 4Ć20.ā COĆox Sample and Waste Tubing
Sample Chamber
CO-ox Pump
Waste Tubing
Reagent Pump
Hemolyzer
FD5
Sample
Connector
Inlet
Outlet
Bubble Trap
2
Menu Code
7