Bayer HealthCare Rapidlab 800 User Manual
Page 255
3Ć83
Maintaining the System
Figure 3Ć58ā.ăFilling a Measurement Sensor
Fill Solution
OĆring
Electrode Compartment
7.
Insert the internal electrode into the electrode compartment and screw it into
place, ensuring that you do not cross-thread the electrode.
8.
Continue with
Reinstalling a Sensor
, page 3-89.
Replacing the Measurement or Sample Ground/
Temperature Sensors
If you need to replace the reference sensor, refer to
Replacing the Reference
Sensor
, page 3-71.
If you need to replace the glucose or lactate biosensors, refer to
Replacing the
Glucose and Lactate Biosensors
, page 3-86.
Materials required:
S appropriate sensor
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
NOTE:
If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
1.
Stop the system from the Menu screen:
a. Select
2
Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
860
2
Menu Code
7