800 Series Operator's Manual
5Ć30
Figure 5Ć16.ā Set Passwords Screen
Type the supervisor
password here.
Type the menu
options password
here.
Retype the menu
options password here.
Retype the supervisor
password here.
NOTE:ĄĂ
You must enter values into both the Supervisor Password and the
Menu Password fields. If you do not use or uniquely define one or the other
password, you can enter the default password. The default for both passwords
is 12345.
3.
Enter the supervisor or menu options password you want to use:
a. Type the password in the Supervisor Password field and press
Enter
.
b. Type the same password in the Supervisor Password Check field and press
Enter
.
You can use the same password for the supervisor and menu passwords or
create a different password for each.
c. Type the password in the Menu Password field and press
Enter
.
d. Type the same password in the Menu Password Check field and press
Enter
.
4.
Press
Done
.
If the Password Check you typed does not match the Password, the system
prompts you to try again. Press
OK
and retype the Password Check.
5.
Define another setup function or press
Exit Menu
to return to the Ready
screen.
If you press Done before you assign passwords, a message appears prompting you
to enter a password. Press
OK
.
You can enter up to 8 alphanumeric characters including the dash in the password
field.