3Ć19

Maintaining the System

Monthly Maintenance

The monthly maintenance schedule is based on the expiration date of the opened
reagents.

Exchanging the Cleaning Solutions

Effective cleaning is accomplished by alternating Cleaning Solution 1 (C1) and
Cleaning Solution 2 (C2) each month.

If . . .

Then remove it and . . .

C1 is on your system

replace it with C2.

C2 is on your system

replace it with C1.

WARNING

Wear safety glasses, gloves, and a laboratory coat when handling the

reagents.

1.

At the Ready screen, press

Enter

.

The Reagent Levels screen appears.

2.

Remove the cleaning solution bottle from the reagent manifold as shown in
Figure 3-7.

Figure 3Ć7.āăRemoving the Cleaning Solution from an 860

Cleaning Solution

3.

Write the date installed in the space provided on the new cleaning solution
bottle.