Bayer HealthCare Rapidlab 800 User Manual
Page 570

EĆ44
800 Series Operator's Manual
Table EĆ38.āă860 Recovery and Precision TestingĊGlucose
*
Level
Mode
n
WRSD
[
Observed
Expected
Recovery
45
Syringe
39
0.9
47
46
102.8
Capillary
30
1.4
46
43
107.9
Microcapillary
30
1.8
44
43
101.2
Microsyringe
36
1.2
45
45
100.6
pH/Lyte
21
1.6
47
47
100.7
81
Syringe
33
2.1
85
84
100.6
Capillary
33
1.4
85
83
102.5
Microcapillary
36
2.5
79
79
99.8
Microsyringe
45
1.9
84
82
102.3
pH/Lyte
24
1.1
81
80
101.1
207
Syringe
33
3.5
205
205
100.4
Capillary
33
2.9
206
203
101.5
Microcapillary
36
3.9
206
205
100.8
Microsyringe
45
3.8
205
203
101.2
pH/Lyte
23
3.1
206
205
100.6
* Performance determined with metabolite recal on. Refer to Glucose Biosensor Calibration in Section 1
for information on performance with recal off.
[
WRSD = within-run standard deviation