Bayer HealthCare Rapidlab 800 User Manual
Page 203

3Ć35
Maintaining the System
5.
Replace the measurement module tubing:
a. Remove the tubing from the connector at the spring-loaded latch in the
measurement block as shown in Figure 3-20.
Figure 3Ć20.āăReplacing the Measurement Module Tubing on an 840
Position 3 Connector
Tubing Groove
Measurement Module Tubing
Connector at the
SpringĆLoaded Latch
Spacer
b. Remove the tubing from the spacer.
c. Disconnect the tubing from the connector at position 3 on the reagent
manifold.
d. Push the new tubing through the spacer.
e. Connect the left end of the tubing to the spring-loaded latch.
f. Connect the right end of the tubing to the connector at position 3.
g. Press the tubing into the tubing groove.
h. Continue with step 7.
6.
Replace the measurement module tubing:
a. Remove the tubing from the connector at the spring-loaded latch in the
measurement block as shown Figure 3-21.
b. Disconnect the tubing from the connector at position 3 on the reagent
manifold.
c. Connect the left end of the new tubing to the spring-loaded latch.
d. Connect the right end of the tubing to the connector at position 3.
e. Press the tubing into the tubing groove.
840
860
850