Bayer HealthCare Rapidlab 800 User Manual
Page 457

System Administration
5Ć47
6.
Press
Yes
.
The QC File Setup screen appears with all data fields empty, as shown in
Figure 5-28.
Figure 5Ć28.ā QC File Setup (Ranges) Screen for an 860
Press Clear Entry to
delete text in a field.
7.
Complete the form.
If you use . . .
Then . . .
the keypad
type the low and high target limits for each parameter. Press
Enter
after you complete each field.
the optional bar
code scanner
scan the target range bar code for each parameter from QC
Expected Values. If you scan the wrong bar code, a message
appears prompting you to scan the correct parameter bar code.
Press
OK
and continue.
When you complete both fields for a parameter, the system automatically
calculates the target value and the action range.
8.
If a second screen of parameters exists, press
Next Screen
and enter values on
that screen.