Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 78
<1. Overview>
1-51
IM 11B08A01-01E
Run
W1
S1
W1
S1
W1
[2E2]
Time
Stream specification 1 (Continuous)
Stop command
Stop
● In the case where a stop command is received during the run mode in the stream
specifi cation status (for N times of measurement) [3E2]
The operation mode is immediately prepared to be changed to the stop mode.
The operation mode changes to the stop mode in the stream specifi cation status (for
N times of measurement) when the measurement of the currently-measured stream is
completed.
In this case, the number of times of measurement is counted again from the beginning.
The stop command can be canceled with a cancel command until the operation mode
changes to the stop mode.
Run
W1
S1(2/8)
W1
S1(1/8)
W1
S1(2/8)
W1
S1(1/8)
W1
[3E2]
Time
Stream specification 1 (8 times)
Stop command
* The number of times of measurement is reset.
Run command
Run command
Stop
Run
● In the case where a stop command is received during the pause mode in the stream
specifi cation status [2E6]
The operation mode is immediately changed to the stop mode in the stream specifi cation
status.
In the stream specifi cation status (for N times of measurement), the number of times of
measurement is counted again from the beginning.
Pause
[2E6]
Time
Stream specification 1
Stop command
Stop
● In the case where a forced stop command is received during the run mode in the stream
specifi cation status [2F2]
The operation mode is immediately changed to the stop mode in the stream specifi cation
status.
In the stream specifi cation status (for N times of measurement), the number of times of
measurement is counted again from the beginning.
2nd Edition : May 11, 2012-00
1