1 a/i status – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 256

<5. EtherLCD>
5-5
IM 11B08A01-01E
5.2.1 A/I
Status
To display the Analog Input (A/I) status, move the cursor (>) with
or
to the A/I Status row
and press
.
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
A/I Status
Slot # 1
AI# Name Value
1 a1 0.000
2 0.000
3 0.000
4 0.000
Menu
Table
Slot#
F1
F2
F3
F4
F5
F6
Figure 5.5
Example of the A/I status screen
F1 (Menu):
Displays the Status Menu screen.
F4 (Table):
Displays the A/I Setup screen (see the 5.4.18 A/I Setting).
F6 (Slot #):
Sets a slot number (applicable only when multiple analog input cards are
inserted).
TIP
• On the initial screen, the smallest slot number among those of the A/I cards inserted is displayed.
• For the slot number setting, only slot numbers of A/I cards can be accepted.
• The Value data are refreshed automatically at a fi xed interval (1 second).
If no A/I cards are inserted, the Status Menu screen displays “Not load” on the bottom.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Slot #
Slot number
1
5
Absolute slot number
AI #
A/I channel number
1
4
Relative channel number
Name
A/I name
8 single-byte alphanumeric
characters
Value
A/I value
-0.25
1.25
If the value stored in the
database is 10000, it is shown
as 1.0000.
2nd Edition : May 11, 2012-00